UP-8000 Multi-Parameter Monitor

UP-8000 Multi-Parameter Monitor
Solicite uma cotação
UP-8000 Multi-Parameter Monitor

Applications:

UP-8000, with the CSM (Cerebral State Monitoring) module and other full-fledged functions, is especially designed for anaesthesia use.

Features:

·12.1" high resolution color LCD screen
·Monitoring standard parameters of ECG, SpO2, RESP, NIBP, TEMP & PR
·CSM module to monitor parameters of CSI, EEG, EMG and BS% with alarm
·Non-invasive monitoring level of consciousness, especially for anaesthesia
·A portable handset with wireless connection to main unit for data processing
·Optimized NIBP modes for adult, pediatric & neonate
·Automatic analysis of 20 arrhythmia waveforms
·S-T segment with automatic & manual analysis
·Digital SpO2 technology with anti-motion & anti-low infusion
·Store & review 400 groups of data & 24-hour ECG waveforms
·Protection against interference from defibrillator & electrosurgical unit
·Built-in thermal printer

Specifications:

CSM

EEG sensitivity

±400μV

Noise level

<2μVp-p, <0.4μV RMS, 1~250Hz

CMR

>140dB

Input impedance

>50Mohm

Sample rate

2000 samples/sec. (14 bits equivalent)

CSI and update

0-100. Filter 6-42Hz, 1sec. update

EMG

0-100 logarithmic. Filter 75-85Hz, 1 sec. update

BS%

0-100%. Filter 2-42Hz, 1 sec. update

Wireless range

Up to 10 meters

ECG

Heart rate display range

15bpm ~ 350bpm

Heart rate display accuracy

± 1% or ± 2bpm, whichever is greater

Sweeping speed

12.5mm /s, 25mm/s, 50mm/s

Sensitivity selection

×1/2, 5mm/mV

×1, 10mm/mV

×2, 20mm/mV

CMRR

≥ 89dB

RESP

Measuring range

0rpm~120rpm

Accuracy

±5% or ±2 rpm

Alarm limits setting range

0rpm~120rpm

TEMP

Measuring range

25.0 ℃ ~45.0℃

Measuring accuracy

±0.2 ℃

Responding time

≤150s

NIBP

Measuring method

Oscillometric Technique

Measuring range

0 mmHg~300mmHg

Accuracy

Maximal mean difference

±5 mmHg

Maximal standard deviation

8 mmHg

Measurement range

Adult

SYS 40 ~ 255 mmHg

MAP 20 ~ 215 mmHg

DIA 10 ~ 195 mmHg

Infant

SYS 40 ~ 200 mmHg

MAP 20 ~ 165 mmHg

DIA 10 ~ 150 mmHg

Neonate

SYS 40 ~ 135 mmHg

MAP 20 ~ 110 mmHg

DIA 10 ~ 95 mmHg

SpO2

Measuring range

35%~100%

Low perfusion capability

0.4%~5%

Accuracy

Not greater than 3% for SpO 2 range from 70% to 100%

PR

Measuring range

30bpm~240bpm

Measuring accuracy

±2bpm or ±2%, whichever is greater

General

Power supply

100~240VAC , 50/60Hz

Rechargeable Li-ion battery

12V 2.3Ah

Display mode

12 inches TFT color LCD

Alarming mode

Audible & visible alarm